PURPOSE:
To test the integrity of the Hepa filter after installation.
PROCEDURE:Implementation ISO 14644-3:2005 standards.
-
Step 1:
- Remove mask of hepa filter box (if any) before performing the test get access to the filter media surface. Step 2:
- The PAO aerosol to be introduced upstream at the negative pressure plenum of equipment, create a challenge aerosol concentration in front of the hepa filter. Step 3:
- Check the upstream challenge aerosol concentration and adjust concentration within limit from 20 to 25 µg/l of air is satisfactory by using the upstream port. Step 4:
-
In upstream position of scan photometer, adjust the concentration upstream from the filter to read 100% value.
-
In downstream position of scan photometer, adjust to read 0% value.
Step 5:
- Test the downstream position and scan the total surface of the filter and the edges of filter frame. Hold the probe approximately 2.5 cm from the filter surface with a scanning speed of not more than 50 mm/s.
ACCEPTANCE CRITERIA:
The downstream challenge aerosol maximum penetration must be < 0.01% of the upstream concentration of challenge aerosol for the test to be accepted.